A scalable high-performance magnetic shield for very long baseline atom interferometry
- verfasst von
- E. Wodey, D. Tell, E. M. Rasel, D. Schlippert, R. Baur, U. Kissling, B. Kölliker, M. Lorenz, M. Marrer, U. Schläpfer, M. Widmer, C. Ufrecht, S. Stuiber, P. Fierlinger
- Abstract
We report on the design, construction, and characterization of a 10 m-long high-performance magnetic shield for very long baseline atom interferometry. We achieve residual fields below 4 nT and longitudinal inhomogeneities below 2.5 nT/m over 8 m along the longitudinal direction. Our modular design can be extended to longer baselines without compromising the shielding performance. Such a setup constrains biases associated with magnetic field gradients to the sub-pm/s2 level in atomic matterwave accelerometry with rubidium atoms and paves the way toward tests of the universality of free fall with atomic test masses beyond the 10-13 level.
- Organisationseinheit(en)
-
Institut für Quantenoptik
QuantumFrontiers
SFB 1227: Designte Quantenzustände der Materie (DQ-mat)
- Externe Organisation(en)
-
IMEDCO AG
Universität Ulm
Technische Universität München (TUM)
- Typ
- Artikel
- Journal
- Review of scientific instruments
- Band
- 91
- ISSN
- 0034-6748
- Publikationsdatum
- 19.03.2020
- Publikationsstatus
- Veröffentlicht
- Peer-reviewed
- Ja
- ASJC Scopus Sachgebiete
- Instrumentierung
- Elektronische Version(en)
-
https://doi.org/10.1063/1.5141340 (Zugang:
Offen)