A scalable high-performance magnetic shield for very long baseline atom interferometry

verfasst von
E. Wodey, D. Tell, E. M. Rasel, D. Schlippert, R. Baur, U. Kissling, B. Kölliker, M. Lorenz, M. Marrer, U. Schläpfer, M. Widmer, C. Ufrecht, S. Stuiber, P. Fierlinger
Abstract

We report on the design, construction, and characterization of a 10 m-long high-performance magnetic shield for very long baseline atom interferometry. We achieve residual fields below 4 nT and longitudinal inhomogeneities below 2.5 nT/m over 8 m along the longitudinal direction. Our modular design can be extended to longer baselines without compromising the shielding performance. Such a setup constrains biases associated with magnetic field gradients to the sub-pm/s2 level in atomic matterwave accelerometry with rubidium atoms and paves the way toward tests of the universality of free fall with atomic test masses beyond the 10-13 level.

Organisationseinheit(en)
Institut für Quantenoptik
QuantumFrontiers
SFB 1227: Designte Quantenzustände der Materie (DQ-mat)
Externe Organisation(en)
IMEDCO AG
Universität Ulm
Technische Universität München (TUM)
Typ
Artikel
Journal
Review of scientific instruments
Band
91
ISSN
0034-6748
Publikationsdatum
19.03.2020
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Instrumentierung
Elektronische Version(en)
https://doi.org/10.1063/1.5141340 (Zugang: Offen)